Hybrid Microscopy Platform Tracks Neuronal Activity and Oxygen Delivery Simultaneously

The brain is a black box of densely packed neurons, blood vessels and immune cells, all locked away in a sealed skull. Washington University in St. Louis’ Song Hu has dedicated his career to unlocking that puzzle box by developing new ways to “see” inside, and his team’s newest innovation could give researchers a new view of how brain cells and blood vessels work together and what happens when the coupling between the two breaks down.

Hu, professor of biomedical engineering at WashU, and his team in the McKelvey School of Engineering, have developed a new way to integrate two powerful techniques: two-photon microscopy and photoacoustic microscopy (TPM-PAM). This will allow researchers to see in real time how the brain vasculature delivers oxygen to neurons. This could have profound implications for studying neurodegenerative disease, stroke, and other diseases connected to neurovascular coupling.

Using this technology gives us a better understanding of these disease mechanisms.

Song Hu, Washington University in St. Louis

Hu and his team shared their results in the journal Nature Communications, including how they combined the two different forms of microscopy into one system and successfully recorded, for the first time, single-neuron calcium activity alongside oxygen release from individual red blood cells in awake mice.

Watching Neuronal Activity and Oxygen Delivery in Action

Two-photon microscopy (TPM) is widely used to image neuronal activity in the brain that is “lit up” by fluorescent probes.

The other technique, photoacoustic microscopy (PAM), uses light-generated sound waves to record blood flow and oxygenation. Brain function is intricately linked to its vasculature. “Thinking” itself consumes energy, but without the view of oxygen delivery that PAM can provide, it’s very difficult to understand any details about how the fuel pipeline works in relation to brain activity.

That’s why it’s important to image both of them, to understand how neuron activity consumes oxygen and how the vasculature changes its dynamics to meet the demand in real time,” Hu said.

With this new technique, Hu and his team were able to bring the two complementary views together at cellular resolution for the first time.

Combining the two for simultaneous imaging of neuronal activity and oxygen delivery, at single-cell level, has never been demonstrated before,” Hu added.

To demonstrate the platform, the team imaged the brains of mice while stimulating their whiskers, tracking neuronal calcium activity alongside changes in blood oxygen delivery. The researchers also used tightly focused laser pulses to block a single capillary or stimulate a single neuron, then followed the resulting changes in nearby neurons and red blood cells. Together, these experiments show how TPM-PAM can combine precise, cell-scale interventions with simultaneous measurements of neuronal activity and oxygen delivery, allowing researchers to move beyond observing correlations and begin probing cause and effect.

Although demonstrated so far only in animal models, the platform could help reveal how brain cells and microvessels coordinate energy use and how that relationship is disrupted in stroke and neurodegenerative disease.

This platform could also inform clinical imaging. Functional MRI does not directly record neurons; it infers changes in brain activity from blood oxygenation and flow. A clearer cellular-scale picture of neurovascular coupling could help researchers interpret these indirect signals much more accurately.

If you can better understand neurovascular coupling, you will make this inference process more accurate,” Hu said.

Making Room for Light and Sound

The technical challenge was solved with help from engineering colleagues at Northwestern University.

Light and sound waves don’t usually play well together.

If you want to integrate the two, you have to make the two forms of energy share the same space, and this is not easy,” Hu said.

Conventional PAM systems use ultrasound detectors that block the optical path or require optical lenses that sacrifice two-photon resolution and light collection. The team’s solution is an optically transparent acoustic sensor built from a polymer micro-ring resonator on a piece of glass.

When ultrasound impinges on the resonator, it slightly deforms the ring and shifts its optical resonance, converting the acoustic signal into a measurable change in light. At the same time, the transparent sensor allows TPM’s excitation and fluorescence light to pass through with minimal interference.

It’s not compromising any of the two imaging technologies,” Hu said.

Source:
Journal reference:

Han, J., et al. (2026). Integrated two-photon and photoacoustic microscopy for single-cell neurometabolic imaging. Nature Communications. DOI:10.1038/s41467-026-75603-7. https://www.nature.com/articles/s41467-026-75603-7.

Comments

The opinions expressed here are the views of the writer and do not necessarily reflect the views and opinions of AZoLifeSciences.
Post a new comment
Post

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.